Spec comparison table
DWX-300LI | DWX-01LI | DWX-05A | ||
Use | Precision test | Ultra-low inductance | General coil component | |
Inductance range | 0.1μH~10mH | 10μH~100mH | ||
Test voltage range | 3~300V | 10~1200V | 5000V | |
Measurement voltage resolution | 0.3V | 1.2V | 100V | |
Maximum applied voltage (standard) | 1μH during 300V | 1μH during 1000V | 10μH during 1000V | |
Waveform sampling resolution | Equivalent to 200MHz | 100MHz | ||
Minimum Test time 1 pulse (wihout screen display) |
High speed modeWithout judgment 10msec With judgment 18msec test modewithout screen display 53msec with screen display 108msec |
200msec | ||
Judgment function | Area Differential Area (Dif Area) Discharge value (Flutter) Breakdown voltage test (BDV) Manual voltage evaluation test (Manual BDV) |
Area Dif. Area (Dif Area) Discharge value (Flutter) Discharge level (Laplacian) |
||
Test Circuit | 4Terminal | 2Terminal | ||
Test cable length range | Recommended 1.5m ( Up to 3m ) | |||
Partial discharge test compatible | Wrong | PD compatible | ||
Additional function | Master waveform storage number 196 models, expansion with CF memory card External control by RS-232C · I / O, test result Log saving · screen hard copy preservation |
|||
External I/O control signal | Impression trigger (Trig)・Application termination (Index) etc. High-speed interface timing support for inspection handlers |
External control such as test start (Start), under test (Busy) | ||
Appearance dimension (Without carry hand) |
345(W)×185(H)×360(D) | |||
Accessories | 4 terminals test cables 1.5m | A Tester Cable (1.65m) | ||
Power supply voltage | AC100/115/220/240 ±5% changeable type |
DWX-300LI
-
- Use
- Precision test
-
- Inductance range
- 0.1μH~10mH
-
- Test voltage range
- 3~300V
-
- Measurement voltage resolution
- 0.3V
-
- Maximum applied voltage (standard)
- 1μH during 300V
-
- Waveform sampling resolution
- Equivalent to 200MHz
-
- Minimum Test time 1 pulse
(wihout screen display) - High speed mode
Without judgment 10msec
With judgment 18msec
test mode
without screen display 53msec
with screen display 108msec
- Minimum Test time 1 pulse
-
- Judgment function
- Area
Differential Area (Dif Area)
Discharge value (Flutter)
Breakdown voltage test (BDV)
Manual voltage evaluation test (Manual BDV)
-
- Test Circuit
- 4 Terminal
-
- Test cable length range
- Recommended 1.5m ( Up to 3m )
-
- Partial discharge test compatible
- Wrong
-
- Additional function
- Master waveform storage number 196 models, expansion with CF memory card
External control by RS-232C · I / O, test result Log saving · screen hard copy preservation
-
- External I/O control signal
- Impression trigger (Trig)・Application termination (Index) etc.
High-speed interface timing support for inspection handlers
-
- Appearance dimension
(Without carry hand) - 345(W)×185(H)×360(D)
- Appearance dimension
-
- Accessories
- 4 terminals test cables 1.5m
-
- Power supply voltage
- AC100/115/220/240 ±5% changeable type
DWX-01LI
-
- Use
- Ultra-low inductance
-
- Inductance range
- 0.1μH~10mH
-
- Test voltage range
- 10~1200V
-
- Measurement voltage resolution
- 1.2V
-
- Maximum applied voltage (standard)
- 1μH during 1000V
-
- Waveform sampling resolution
- Equivalent to 200MHz
-
- Minimum Test time 1 pulse
(wihout screen display) - High speed mode
Without judgment 10msec
With judgment 18msec
test mode
without screen display 53msec
with screen display 108msec
- Minimum Test time 1 pulse
-
- Judgment function
- Area
Differential Area (Dif Area)
Discharge value (Flutter)
Breakdown voltage test (BDV)
Manual voltage evaluation test (Manual BDV)
-
- Test Circuit
- 4 Terminal
-
- Test cable length range
- Recommended 1.5m ( Up to 3m )
-
- Partial discharge test compatible
- Wrong
-
- Additional function
- Master waveform storage number 196 models, expansion with CF memory card
External control by RS-232C · I / O, test result Log saving · screen hard copy preservation
-
- External I/O control signal
- Impression trigger (Trig)・Application termination (Index) etc.
High-speed interface timing support for inspection handlers
-
- Appearance dimension
(Without carry hand) - 345(W)×185(H)×360(D)
- Appearance dimension
-
- Accessories
- 4 terminals test cables 1.5m
-
- Power supply voltage
- AC100/115/220/240 ±5% changeable type
DWX-05A
-
- Use
- General coil component
-
- Inductance range
- 10μH~100mH
-
- Test voltage range
- 5000V
-
- Measurement voltage resolution
- 100V
-
- Maximum applied voltage (standard)
- 10μH during 1000V
-
- Waveform sampling resolution
- 100MHz
-
- Minimum Test time 1 pulse
(wihout screen display) - 200msec
- Minimum Test time 1 pulse
-
- Judgment function
- Area
Dif. Area (Dif Area)
Discharge value (Flutter)
Discharge level (Laplacian)
-
- Test Circuit
- 2 Terminal
-
- Test cable length range
- Recommended 1.5m ( Up to 3m )
-
- Partial discharge test compatible
- PD compatible
-
- Additional function
- Master waveform storage number 196 models, expansion with CF memory card
External control by RS-232C · I / O, test result Log saving · screen hard copy preservation
-
- External I/O control signal
- External control such as test start (Start), under test (Busy)
-
- Appearance dimension
(Without carry hand) - 345(W)×185(H)×360(D)
- Appearance dimension
-
- Accessories
- A Tester Cable (1.65m)
-
- Power supply voltage
- AC100/115/220/240 ±5% changeable type